Mô tả sản phẩm: Máy sốc tim Philips HeartStart FR3 AED
Models |
861388 (text display) |
Waveform |
SMART biphasic Truncated Exponential waveform parameters adjust as a function of patient impedance. Adult nominal peak current 32A (150J into a 50 ohm load); pediatric nominal peak current 19A (50J into a 50 ohm load) using optional Infant/Child Key |
Shock delivery |
Via defibrillator pads placed in the anterior-anterior (Lead II) position for adults; anterior-posterior position for infants and children under 55 lbs (25 kg) or 8 years old |
Controls |
On/Off button, shock button, option buttons. Auto-On feature, when used with the optional FR3 carry case, enables FR3 to power up when case lid is opened |
Indicators |
High-resolution color LCD, beeper, voice prompts, tones and chirps, audio speaker, connector socket, ready light, shock button |
Advanced Mode |
Configurable using optional HeartStart Configure software |
ECG Display
Screen |
LCD color display, 320 x 240 pixels. 2.8" x 2.1" (7.2 cm x 5.4 cm) |
Bandwidth |
1 Hz to 30 Hz (-3dB), nominal (non-diagnostic) |
Monitored Lead |
Lead II using anterior-anterior adult pads placement |
Physical
Size |
2.7" high x 5.3" wide x 8.7" deep (6.9 cm x 13.5 cm x 22.1 cm) |
Weight |
3 lbs 8 oz (1.6 kg) with FR3 standard battery installed |
Environmental/physical requirements
Sealing |
Meets IEC529 class IP55 with battery installed |
Temperature |
Operating/standby: 32°–122°F (0°– 50°C) |
Altitude |
0 –15,000 feet (0 –4,572 meters) |
Shock/drop |
Meets MIL-STD-810F 516.4, Procedure IV |
Abuse tolerance |
After a ibe-meter drop to any edge, corner, or surface in standby mode |
Vibration |
Meets MIL-STD-810F 514.5 C-17 |
Patient Impedance Range |
Minimum: 15 ohm (internal defibrillation); 25 ohm (external defibrillation) |
AED Mode |
Shock advisory sensitivity and specificity meet AAMI DF-39 guidelines |
Bluetooth 2.0 Class II Wireless Transceiver Module (optional)
Function |
Transmit retrospective event data or configuration setting wirelessly |
Patient analysis system
ECG analysis |
Evaluates impedance of defibrillator pads for proper contact with patient skin, evaluates the ECG rhythm and signal quality to determine if a shock is appropriate; also detects artifact and pacemaker |
SMART CPR |
Evaluates key characteristics of the presenting VF and determines the initial therapy: shock first, or CPR first quickly followed by a shock |
Sensitivity/specificity |
Meets AAMI DF80 requirements and AHA recommendations for adult defibrillation |
Quick Shock |
Typically arms in <8 seconds from the end of the "Stop CPR" prompt |
FR3 primary battery
Type |
12 VDC, 4.7 Ah, lithium manganese dioxide long-life primary cells |
Capacity |
Typically 300 shocks or 12 hours of operating time at 77° F (25° C) when configured for monitoring after No Shock Advised (NSA) |
Standby life |
3 years minimum when stored under standby environmental conditions (battery installed) |
Shelf life |
5 years |
SMART Pads III
Applications |
Disposable, multifunction defibrillation pads for adult or infant/child patients. Time-saving peel and place pads can be removed from packaging and stored in the FR3 carry case. Pads can be preconnected to FR3, which enables testing during FR3's routine self-test. |
Infant/Child Key (optional)
Function |
Selects therapy for infants or children under 55 lbs (25 kg) or 8 years old |
FR3 data card (optional)
Function |
Stores a minimum of 8 hours of ECG, event, and if configured, voice recording. Can also be used for configuring FR3. |
Automated and user-activated self-tests
Automatic self-tests |
Test internal circuitry, waveform delivery system, ECG acquisition, temperature, status (or readiness) of attached accessories (SMART Pads III and FR3 data card) and battery |
Automated self-test frequency |
Daily, weekly, monthly, power on, and runtime during all modes of operation |
User initiated tests |
Automatic self-tests plus tone, display, and button performance |
FR3 training battery and training pads (optional)
Function |
Places FR3 into a scenario-based training mode and simulates shock therapy |
Type |
10.8 Volt, 4.5 Ah Li-ion battery |