Mô tả sản phẩm: Máy sốc tim Philips Efficia DFM100 Defibrillator/Monitor
General
Approximate Dimensions:
|
23.5 cm (H) x 29 cm (W) x 20.5 cm (D);
|
|
Approximate Weight (without battery):
|
5.66 kg; 12.5 lbs
|
|
Standard Operator Position:
|
Within one meter (3 feet) of the device
|
|
Power:
|
Rechargeable Lithium Ion battery; AC power using a protectively grounded outlet
|
|
Defibrillator
Waveform:
|
Biphasic Truncated Exponential; waveform parameters adjusted as a function of patient impedance
|
|
Shock Delivery:
|
Via multifunction electrode pads or paddles
|
|
Shock Series:
|
Configurable energy escalation in a series Leads Off Sensing and PCI
|
|
Sensing for Pads/Paddles:
|
Apply 500 nA rms (571 Hz); 200 uA rms (32 KHz)
|
|
Charge times:
|
|
|
Patient Impedance Range:
|
Minimum: 25 ohm (external defibrillation); 15 ohm (internal defibrillation) Maximum: 250 ohm. Actual functional range may exceed these values
|
|
Display
Size:
|
Approximately 7 in (17.8 cm) diagonal viewing area
|
|
Type:
|
Color TFT LCD
|
|
Resolution:
|
800 x 480 pixels (VGA) with 32 brightness levels per color
|
|
Sweep Speed:
|
25 mm/s ± 10% nominal (stationary trace; sweeping erase bar) for ECG and SpO2; capnogram wave is 6.25 mm/s ± 10%
|
|
Wave Viewing Time:
|
6.5 sec ± 10%
|
|
Battery
Type:
|
Rechargeable, Lithium Ion; See battery label for capacity information
|
|
Approximate Dimensions:
|
28.5 mm (H) x 80 mm (W) x 145.7 mm (L); 1.1 in (H) x 3.1 in (W) x 5.7 in (L)
|
|
Approximate Weight:
|
Approximately 0.44 kg (1 lb)
|
|
Capacity:
|
With a new fully charged battery, at 20 °C (68 °F), one of the following:
|
|
Battery Indicators:
|
Battery gauge on battery, capacity indicator on display, power indicators on front of device; flashing RFU indicator, audio beep and Low Battery messages on the display for low battery condition. When a low battery message first appears there is still enough energy for at least 10 minutes of monitoring and 6 maximum energy discharges.
|
|
Thermal Array Printer
Continuous ECG Strip:
|
The Print key starts and stops the strip. The printer can be configured to be run real time or with a 10 second delay. The strip prints the primary ECG lead and a second wave with event annotations and measurements.
|
|
Auto Printing:
|
The printer can be configured to automatically print on Mark Events, Charge, Shock and Alarm.
|
|
Reports:
|
The following can be printed:
|
|
Speed:
|
25 mm/s with an accuracy of ± 5 %
|
|
Amplitude Accuracy:
|
5 % for offset voltages of ± 300 mV at 5 Hz
|
|
Paper Size:
|
50 mm (W) x 20 m (L)
|
|
Environmental
Temperature:
|
Operating temperature for the device: 0 °C to 45 °C (32 °F to 113 °F); Operating temperature range for EtCO2: 0 °C to 40 °C (32 °F to 104 °F); Storage temperature range for the device without battery: -20 °C to 70 °C (-4 °F to 158 °F)
|
|
Humidity:
|
Up to 95 % relative humidity
|
|
Atmospheric Range:
|
Operating and Storage – 1014 mbar to 572 mbar (0 to 15,000 ft.; 0 to 4,500 m)
|
|
Shock:
|
Operating: Half-sine waveform, duration ≤ 11 ms, acceleration ≥ 15.3 G, 3 shocks per face Non-operating: Trapezoidal waveform, acceleration 30 G, velocity change 7.42 m/s ± 10 % 1 shock per face
|
|
Water/Solid Ingress Resistance:
|
Meets Ingress Protection level IP44
|
|
EMC:
|
Complies with the requirements of standard EN 60601-1-2:2002
|
|
Safety:
|
Meets EN 60601-2-4:2003, EN 60601-1:1990
|
|
Mode of Operation:
|
Continuous
|
|
AC Line Powered:
|
100 to 240 VAC, 50 or 60 Hz, 1 to 0.46 A, Class I Equipment
|
|
Battery Powered:
|
Minimum 14.4 V, Rechargeable Lithium Ion |