Mô tả sản phẩm: Máy sốc tim Philips HeartStart XL+ Defibrillator/Monitor
Physical Dimensions
Size |
9˝ high x 11.6˝ wide x 10.9˝ deep (23 cm x 29.6 cm x 27.9 cm) |
Weight |
14.7 lbs (6.6 kg), includes one battery, one new roll of paper, one pads cable. Incremental weight of external standard paddles and paddle tray is less than 3 lbs (1.3 kg) |
Standard operator position | Within one meter (3 feet) of the device |
Power supply
Type | Rechargeable lithium ion battery; AC power using a protectively grounded outlet |
Defibrillator
Waveform | Biphasic Truncated Exponential. Waveform parameters adjusted as a function of patient impedance |
Shock delivery | Via multifunction electrode pads or paddles |
Shock series | Configurable energy escalation in a series |
Leads off sensing and PCI sensing for pads/paddles | Apply 500nA rms (571Hz); 200uA rms (32KHz) |
Charge time |
|
Patient impedance range |
Minimum: 25 ohm (external defibrillation);15 ohm (internal defibrillation); Maximum: 250 ohm. Actual functional range may exceed these values |
Printer
Continuous ECG strip | The Print button starts and stops the strip. The printer can be configured to be run real time or with a 10-second delay. The strip prints the primary ECG lead and a second wave with event annotations and measurements |
Auto printing | The printer can be configured to automatically print on Mark Events, Charge, Shock and Alarm |
Reports | The following can be printed: Event Summary (Long or Short), Vital Signs Trends, Operational Check, Configuration, Status Log, Device Information |
Speed | 25 mm/s with an accuracy of +/-5% |
Amplitude accuracy | 5% for offset voltages of +/- 300 mV at 5Hz |
Paper size | 50 mm wide x 30 m long |
Battery
Type | Rechargeable, lithium ion; see battery label for capacity information |
Dimensions | 1˝ high x 4.5˝ wide x 5.7˝ long (23.6 mm high x 116 mm wide x 146 mm long) |
Weight | Approximately 1.5 lbs (.68kg) |
Charge time, device off | With the temperature between 0-35° C (32-95° F), less than 3 hours to 100% capacity; less than 2 hours to 80% capacity |
Life | Approximately 3 years |
Capacity | With a new fully charged battery, at 20 °C (68 °F), one of the following: At least 3 hours of monitoring (ECG and SpO2 monitored continuously and NBP sampled every 15 minutes) followed by 20 full-energy charge/shocks; OR at least two hours of pacing (180ppm at 140mA with 40 msec pulse width) while monitoring (ECG and SpO2 monitored continuously and NBP sampled every 15 minutes) followed by 20 full-energy charge/shocks; OR at least 175 full energy charge/shocks |
Battery indicators | Battery gauge on battery, capacity indicator on display, power indicators on front of device; flashing RFU indicator, chirp and Low Battery messages on the display for low battery condition. When a low battery message first appears there is still enough energy for at least 10 minutes of monitoring and 6 maximum energy discharges |
Battery storage | Storing the battery for extended periods at temperatures above 40° C (104° F) reduces battery capacity and degrades battery life |
Display
Size | Approximately 6.5 in (16.5 cm) diagonal viewing area |
Type | Approximately 6.5 in (16.5 cm) diagonal viewing area |
Resolution | 640 x 480 pixels (VGA) with 32 brightness levels per color |
Sweep speed | 20 mm/s nominal (stationary trace; sweeping erase bar) for ECG and SpO2 |
Wave viewing time | 5.2 sec |
Environmental
Temperature |
0°C to 45°C (32°F to 113°F) operating; -20°C to 70°C (-4°F to 158°F) storage |
Humidity | Up to 95% relative humidity |
Atmospheric pressure range | Operating and storage - 1014 mbar to 572 mbar(0 to 15,000 ft; 0 to 4,500 m) |
Shock – operating | Half-sine waveform, duration ≤11 ms, acceleration ≥ 15.3 G, 3 shocks per face |
Shock – non-operating | Trapezoidal waveform, acceleration 30G, velocity change 7.42 m/s +/-10% 1 shock per face |
Water/solids ingress resistance | Meets Ingress Protection level IP21 |
EMC | Complies with the requirements of standard EN 60601-1-2:2002 |
Safety | Meets UL 60601-1 (1st edition), EN 60601-2-4:2003, EN 60601-1:1990 |
Mode of operation | Continuous |
AC Line powered | 100-240 VAC, 50 or 60 Hz, 1-0.46A, Class I Equipment |
Battery powered | Minimum 14.4 V, rechargeable lithium ion |