Test Specifications
|
||
The accuracy specification for the display range is defined as ± (%reading + digit counts) at 23 °C ± 5 °C, ≤ 75 % RH. Between 0 °C and 18 °C and between 28 °C and 40 °C, accuracy specifications may degrade by 0.1 x (accuracy specification) per °C. | ||
The measurement range meets the service operating errors specified in EN61557-1: 1997, EN61557-2: 1997 , EN61557-4: 1997. | ||
Power-on Test | The test indicates reversed L-N, missing PE, and measures the mains voltage and frequency. | |
Display range: | 90 V to 264 V | |
Accuracy at 50 Hz: | ± (2% + 3 counts) | |
Resolution: | 0.1 V (1 V - model 6200) | |
Input impedance | > 1 MΩ // 2.2 nF | |
Maximum input mains voltage: | 300 V | |
Earth Bond Test (Rpe) | Display range: | 0 to 19.99 Ω |
Accuracy (after Bond Test zeroing): | ± (2.5% + 4 counts) | |
Resolution: | 0.01 Ω | |
Test current: | 200 mA AC -0% +40% into 1.99 Ω | |
25 A AC ± 20% into 25 mΩ at 230 V | ||
Open circuit voltage: | > 4 V AC, < 24 V AC | |
Bond test zeroing: | can subtract up to 1.99 Ω | |
Insulation Test (Riso) | Display range: | 0 to 299 MΩ |
Accuracy: | ± (5% + 2 counts) from 0.1 to 300 MΩ | |
Resolution: | 0.01 MΩ (0 to 19.99 MΩ) | |
0.1 MΩ (20 to 199.9 MΩ) | ||
1 MΩ (200 to 299 MΩ) | ||
Test voltage: | 500 V DC -0% + 10% at 500 kΩ load | |
Test current: | > 1 mA at 500 kΩ load, < 15 mA at 0 Ω | |
Auto discharge time: | < 0.5 s for 1 μF | |
Max. capacitive load: | operational up to 1 μF | |
Touch Current Test | Display range: | 0 to 1.99 mA AC |
Accuracy: | ± (4% + 2 counts) | |
Resolution: | 0.01 mA | |
Internal resistance (via probe): | 2 kΩ | |
Measuring method: | probe | |
The appliance under test is energized at mains potential. | ||
Substitute Leakage Current Test | Display range: | 0 to 19.99 mA AC |
Accuracy: | ± (2.5% + 3 counts) | |
Resolution: | 0.01 mA | |
Test voltage: | 35 V AC ± 20% | |
Load/Leakage Test: Load Current | Display range: | 0 to 13 A |
Accuracy: | ± (4% + 2 counts) | |
Resolution: | 0.1 A | |
The appliance under test is energized at mains potential. | ||
Load/Leakage Test: Load Power | Display range: | 0 to 999 VA |
1.0 kVA to 3.2 kVA | ||
Accuracy: | ± (5% + 3 counts) | |
Resolution: | 1 VA (0 to 999 VA) | |
0.1 kVA (1.0 kVA to 3.2 kVA) | ||
The appliance under test is energized at mains potential. | ||
Load/Leakage Test: Leakage Current | Display range: | 0 to 19.99 mA |
Accuracy: | ± (4% + 4 counts) | |
Resolution: | 0.01 mA | |
The appliance under test is energized at mains potential. | ||
PELV Test | Accuracy at 50 Hz: | ± (2% + 3 counts) |
Overload protection: | 300 Vrms | |
Warning threshold: | 25 Vrms | |
Environmental Specifications | ||
Operating Temperature | 0 to 40 °C | |
Relative Humidity | non condensing < 10 °C | |
95% from 10 to 30 °C | ||
75% from 30 to 40 °C | ||
Safety Specifications | ||
Safety Rating | EN61010-1 CAT II, 300 V | |
Mechanical & General Specifications | ||
Size | 200 mm (L) x 275 mm (W) x 100 mm (H) | |
Weight | 3 kg | |
Power Supply | V + 10 % - 15 %, 50 Hz ± 2 % | |
Power Consumption (Tester) | 13 W typical (idle) | |
60 W max. during 25A Bond Test | ||
Storage | Temperature: | -10 to 60 °C |
Corrosion: | 70 °C @ 95 % RH for 5 days max. | |
Operating Altitude | 0 up to 2000 m | |
Sealing | IP-40 (enclosure), IP-20 (connectors) | |
EMC | complies with EN61326-1, criteria B | |
EMI Immunity | 3 V/m |