Mô tả sản phẩm: Fluke 123/S ScopeMeter
Specifications
|
||
Oscilloscope Bandwidth | Fluke 125: | 40 MHz |
Fluke 124: | 40 MHz | |
Fluke 123: | 20 MHz | |
Dual input True RMS Meter | Vdc, Vac, Vac+Vdc, Ohms,Continuity , Diode-test | |
Current, °C, °F, Capacitance, dBV, dBM, Crest Factor, Touch Hold and Zeroset | ||
TrendPlot Recording | Automatic vertical scaling and time compression. | |
Display the actual, MIN, MAX and AVG reading. | ||
Cursors | Fluke 125/124: | ∆T, 1/∆T, V, ∆V, rise- and fall time |
Real-time sample rate | 25 MS/s | |
Equivalent sampling rate | up to 2.5 GS/s (125, 124) or 1.25 GS/s (123). | |
Inputs and digitizers | 2 | |
Timebase range | Fluke 125/124: | 20 ns - 1 min/div |
Fluke 123: | 10 ns - 1 min/div | |
Input sensitivity | 5 mV - 500 V/div | |
Trigger types | Connect-and-View™, Free Run, Single Shot, Edge, Video, External | |
Glitch capture | 40 ns | |
Scope measurements | Automatic measurements: 26 (all instruments) On top, Fluke 125 has additional measurement functions for Power (W), VA, VAR and Power Factor (PF) for single phase and balanced-three-phase (delta) power systems. | |
Harmonics mode | Fluke 125 has capbility to analyze the harmonics contents of power signals. See below for details. | |
Maximum record length | 512 min/max pts per input | |
Memory | Fluke 125/124: | 20 screens and setups |
Fluke 123: | 10 screens and setups | |
Additional Power measurement capabilities (Fluke 125 only) | ||
Measurement types: | • | Watt, |
• | VA, | |
• | VAR, | |
• | Power Factor (PF) | |
Power configuration: | Single phase or balanced three-phase (delta-configuration) mains supply | |
Voltage measurement | Channel A; using STL120, voltage probe or direct input | |
Current measurement | Channel B, using i400s (included) or other compatible current clamp | |
Current Clamp or Shunt sensitivity: | 0.1 / 1 / 10 / 100 / 1000 mV/A, 10 mV/mA and 400 mV/A. | |
Harmonics Mode (Fluke 125 only) | ||
Converts waveform information into a harmonics display (using FFT processing) which shows the relative amplitudes of the 1st up to the 33rd harmonic | ||
Analyzed waveform: | Voltage waveform(Ch.A), Current wavefrom (Ch.B) or Power (Ch.A x Ch.B), automatically generated. | |
Harmonics Frequency range: | DC up to 33rd harmonic (for fundamental up to 60 Hz); | |
DC up to 24th harmonic (for fundamental up to 400 Hz). | ||
Display | Bargraph showing 1st up to 33rd harmonic and DC; amplitude displayed in % relative to fundamental | |
Measurements: | Relative amplitude of individual harmonics; | |
THD in %r or %f. | ||
Environmental Specifications | ||
Operating Temperature | 0 ºC to 50 ºC | |
Storage Temperature | -20 °C to +60 °C | |
Operating Altitude | 3,000 m | |
Safety Specifications | ||
Electrical Safety | 600V CAT III (EN61010-1) | |
Mechanical & General Specifications | ||
RS232/USB Interface | Optically isolated via optional PM9080 or OC4USB for interfacing to Printer or PC |