ECG
|
HR range
|
0-300bpm
|
Resolution
|
1bpm
|
Accuracy
|
?% or ?bpm, Choose the higher one
|
Bandwidth for ECG
|
0.5Hz-75Hz Monitoring mode: 0.5Hz-40Hz; Diagnosis mode: 0.5Hz-75Hz
|
Lead selection
|
standard I, II, III; aVR, aVL, aVF; V1-V6
|
ST segment
|
-2.0-2.0mV
|
Arrhythmia analysis
|
Yes |
Lead-off alarm
|
Audible and visual |
Sweep speed |
12.5, 25, 50mm/sec |
Gain selection |
X1/2, x1, x2, x4 and Auto |
Protection |
Prevent AC4000V/50Hz voltage in isolation and work against electrosurgical interference, AC, EMG and defibrillation |
Alarm |
High/Low limits, automatic saving |
NIBP
|
Method
|
Intelligent oscillation
|
Parameters
|
Syst/Dias/Mean, Pulse
|
Operation modes
|
Manual/Auto/STAT (5 minutes)
|
Pressure unit
|
mmHg/kPa
|
Measurement range |
Adult: 10-270mmHg Child: 10-200mmHg Neonate:10-135mmHg |
Range of cuff pressure |
0-300mmHg |
Cuff inflation
|
<10 secs
|
Cuff deflation |
<20 secs |
Alarm |
High/Low limits, automatic saving |
RESP
|
Method
|
Chest impedance |
Respiration range |
0-150bpm |
Accuracy |
?bpm |
Alarm |
High/Low limits, automatic saving |
TEMP
|
Range
|
0℃-50℃ |
Resolution |
0.1℃ |
Accuracy |
?.2℃ |
Probe |
Standard: Permanent dermal temperature probe Optional: Esophageal or rectal temperature probe |
Alarm |
High/Low limits, automatic saving |
SpO2
|
Display |
SpO2 value, pulse histogram, waveform, pulse |
Range of SpO2 |
0%-100% adult/child/neonate |
Resolution |
1% |
Accuracy |
?% (75%-99%) ?% (50%-74%) ?% ( 40%-49%) |
Range of pulse rate |
30~240bpm |
Accuracy of pulse rate |
?bpm |
Sensor |
Standard: finger tip Optional: infant Y-type clip, neonate wrap |
Alarm |
High/Low limits, automatic saving |
CO2
|
Measurement mode |
Side-stream |
Parameters |
Respiration rate, EtCO2, InsCO2 |
Resolution |
?mmHg |
Alarm range |
0mmHg-120mmHg |
EtCO2 alarm default setting |
42/20mmHg |
InsCO2 alarm default setting |
10/0mmHg |
Sampling rate |
60 times/second |
Temperature for use |
10℃-40℃ |
Temperature for storage |
0℃-65℃ |
Relative moisture for use and storage |
5%-95% |
Atmospheric pressure range |
500mmHg-800mmHg |
IBP
|
Measurement mode |
IBP transducer Transducer sensitivity: 5μV/V/mmHg |
Measurement range |
-30mmHg-300mmHg |
Parameters |
SYST, DIAS, MEAN |
Calibration mode |
Zero and 100mmHg calibration (optional) |
Resolution |
0-100mmHg ?mmHg |
Measurement sites |
100mmHg-300mmHg ?%ART, PA, CVP, RAP, LAP, ICP |
Sampling rate |
512 times/second |
Temperature for storage |
0℃-65℃ |
Recorder |
Recorder type |
Built-in thermal recorder |
Recording mode |
Recording waveform, text in real time |
Configuration
|
Display
|
12.1″ color TFT display
|
Standard |
Three-channel ECG, NIBP, SpO2, RESP, TEMP, Pulse Rate, Built-in battery |
Optional
|
IBP, CO2 (mainstream or sidestream), built-in thermal recorder
|
Others
|
Power supply |
AC100-250V, 50/60Hz |
Power consumption |
≤80VA |
Battery |
Rechargeable sealed lead acid battery |
Operating time |
2 hours or more |
Networking |
Standard RS232 |
Trend time |
1-96 hours |
Net/Gross weight |
6.5kg/11.5kg |
Packaging |
500L X 320WX 460H mm |
Safety standard |
IEC601-1 Class II, Type BF, CF |